Impedance (Electrode Site)

Hardware
Impedance is the opposition an electrode site presents to current at a given frequency, measured across the electrode–tissue interface in ohms and typically quoted, by convention, at 1 kHz.

The electrode–tissue interface is not purely resistive. For metal sites, charge crosses the interface through a double layer that behaves largely as a capacitor, so the impedance depends on the area of the site, its metal, its surface texture, and any coating. Site impedance also depends on frequency, and it decreases as frequency increases.

Impedance affects the recorded data in two ways. First, the interface generates thermal noise in proportion to the square root of its resistance, so a site with higher impedance has a higher noise floor before the amplifier adds its own noise. Second, the site impedance forms a voltage divider with the input impedance of the amplifier. When the two are comparable, part of the signal is lost, so the amplifier input impedance should be well above the site impedance. For example, the Intan RHD chips in NeuroNexus headstages present 13 MΩ at 1 kHz, more than ten times the impedance of a 1 MΩ site. The typical range follows from site area. A 177 µm² site on a NeuroNexus probe typically measures 200 kΩ to 1 MΩ at 1 kHz. Larger sites measure lower, and smaller sites measure higher. Impedance can also be lowered without increasing the site area. Z-Coat increases the effective surface area within the same footprint and lowers impedance by roughly an order of magnitude, and SIROF sites measure nominally below 100 kΩ at 177 µm² (Cogan (2008)).

Impedance also limits stimulation, because it limits the charge that a site can pass before the interface voltage reaches the level at which electrolysis of water begins. Impedance is also a primary diagnostic measurement. A site that measures far above its typical value may have a broken trace, a contaminated surface, or a dried residue, and a site that measures far below may have a short. Measuring every site before a session is the first check in probe setup. The headstage chip measures impedance in place, so the check is run from the acquisition software.

Typical values

Measurement frequency, by convention1 kHz
Typical impedance, 177 µm² site200 kΩ to 1 MΩ
Amplifier input impedance, working ruleat least 10 × site impedance
Thermal noisescales with the square root of interface resistance

Cite this term

NeuroNexus. "Impedance (Electrode Site)." NeuroNexus Glossary, updated 2026-11-10. https://www.neuronexus.com/resources/glossary/impedance


Category:

Hardware

· Physical probe specifications, electrode properties, and hardware concepts

Ask

Cart

History

Resources

Describe your setup and we'll match you to the right probe

Products, compatibility, documents and how to choose. Quotes and orders go to our team.

Oaks 0.8α

Ctrl+K

Save products while browsing — they'll appear here

Look for the "Add to cart" button on product pages

Log in to see your conversation history and quotes

Sign in

Featured resources

Contextual documents and guides will appear here based on the current page.