NeuroNexus · neuronexus.com · Glossary
Shank Separation
Shank separation is the center-to-center distance between neighboring shanks on a multi-shank silicon probe, in micrometers. It determines, in part, whether the shanks sample one structure at several points or separate structures at once.
It is the fourth field of a multi-shank probe name (the 200 in A4x8-tet-5mm-100-200-177), and it is the across-shank counterpart of site pitch.
The choice of separation follows from the detection radius. Spikes can typically be assigned to single units within about 50 to 100 µm of a site (Quian Quiroga & Panzeri (2009)), and sites 200 µm apart on a silicon probe recorded separate units (Drake et al. (1988)). Shanks separated by 200 µm therefore generally sample adjacent columns of tissue with little overlap. This is the separation on most NeuroNexus catalog multi-shank designs. A closer separation (125 or 150 µm) allows neighboring shanks to record some of the same neurons, and it is used where unit isolation across shanks is important. A wider separation (400 to 500 µm) places shanks in different structures or different cortical areas in one insertion. The Matrix platform reaches 1.5 mm and more to sample a volume. Shank separation and shank count together determine the width of the site span.
Typical values
| Catalog multi-shank designs | 125 to 500 µm; 200 µm on about half |
|---|---|
| Matrix 3D platform | up to 1.8 mm |
| Design-space parameter | Shank separation (shankSepUm) |
References
- Quian Quiroga & Panzeri, 2009 — Extracting information from neuronal populations: information theory and decoding approaches. Nat Rev Neurosci 10(3), 173–185
- Drake et al., 1988 — Performance of planar multisite microprobes in recording extracellular single-unit intracortical activity. IEEE Trans Biomed Eng 35(9), 719–732
Related Terms
Cite this term
NeuroNexus. "Shank Separation." NeuroNexus Glossary, updated 2026-11-10. https://www.neuronexus.com/resources/glossary/shank-separation