High-Definition Deep Brain Stimulation (HD-DBS) with rDBSA
Precision targeting, electrode configurations and implementation strategy for high-definition DBS research with the NeuroNexus rDBSA array.
6 min read
Updated October 12, 2026
The rDBSA ↗ (research Deep Brain Stimulation Array) brings clinical DBS electrode geometry into research use. It combines microstimulation and recording on one array, acute or chronic, and reaches targets a standard laminar probe cannot.
Research use only. NeuroNexus products and the procedures described here are for non-clinical research use, and must not be used in human or veterinary medical procedures.
What the rDBSA is#
Clinical DBS leads are built for therapy, so their contacts are large and few. A research array asks a different question: where, exactly, does the effect come from. The rDBSA answers it by placing more and smaller contacts along the same trajectory a clinical lead would take.
The result is a linear contact array on a long, slim shank, addressable channel by channel. Contacts can stimulate, record, or do both within one session.
Recording and stimulation on one array#
Separating the stimulating contact from the recording contact by a known distance is the experimental point. It lets you ask what the tissue did, at a measured offset from where the current went in.
Two consequences follow for experimental design:
- Stimulation artifact sets your usable window. Plan the recovery interval before the protocol, not after the first session.
- Contact selection is a variable, not a setting. Record the exact contacts used in every trial; the spatial claim depends on it.
Closed-loop protocols use the same property. Recording contacts supply the trigger, stimulating contacts deliver the response, and both live on one implanted array.
Deep targets and long trajectories#
Subthalamic nucleus and globus pallidus work in large-brain models needs shank lengths a cortical probe does not provide. rDBSA designs run to tens of millimeters of penetration depth, on a lead diameter chosen to keep the insertion tract narrow.
Two things follow from the length:
- Trajectory error accumulates with depth. A one-degree angular error is small at 5 mm and material at 45 mm. Plan the approach with imaging, and verify the track histologically.
- Insertion speed matters more, not less. A long shank displaces tissue along its whole path.
Small-animal configurations#
A rat-sized rDBSA exists for rodent DBS models, developed with a collaborating rodent DBS laboratory. It keeps the contact arrangement and scales the geometry to rat anatomy. A protocol developed in rodents then transfers to a larger model without changing the recording approach.
Research applications#
- Movement disorder models — stimulation of STN and GP, with simultaneous recording of the stimulated and downstream structures.
- Psychiatric and pain models — targeted stimulation where the mechanism, not the therapy, is the object of study.
- Epilepsy models — responsive stimulation triggered from recorded activity.
- Mechanism studies — separating direct activation from network-level effect by recording at a known distance from the stimulating contact.
Experimental considerations#
| Consideration | What goes wrong | Settle it |
|---|---|---|
| Stereotactic compatibility | The holder does not grip the lead diameter you ordered, discovered on surgery day | Confirm the manipulator's collet against the ordered lead diameter before the first surgery |
| Chronic fixation | A 45-60 mm lead is a lever; torque at the entry point unseats the anchor | Anchor close to the entry point and support the span between skull and connector |
| Charge density | Small contacts reach charge-density limits at currents a clinical contact tolerates | Establish the safe envelope for your contact geometry before the protocol, not during it |
| Impedance tracking | A mechanical failure is read as a biological change | Measure at implant and every session; a rise on one contact is usually mechanical |
Configurations and ordering#
rDBSA is configured to the target. Eighteen package configurations are currently orderable: the 32-contact lead on an Omnetics 32-pin connector and the 40-contact lead on an MRI Omnetics 42-pin connector, each in three housings and three cannula lengths:
| Configuration | Channels | Cannula length | Protective covering | Connector |
|---|---|---|---|---|
| rdbsa32_a_cl35 | 32 | 35 mm | 3D-printed polymer case | Omnetics 32-pin |
| rdbsa32_c_cl35 | 32 | 35 mm | 3D-printed titanium housing | Omnetics 32-pin |
| rdbsa32_cc_cl35 | 32 | 35 mm | Epoxy | Omnetics 32-pin |
| rdbsa32_a_cl45 | 32 | 45 mm | 3D-printed polymer case | Omnetics 32-pin |
| rdbsa32_c_cl45 | 32 | 45 mm | 3D-printed titanium housing | Omnetics 32-pin |
| rdbsa32_cc_cl45 | 32 | 45 mm | Epoxy | Omnetics 32-pin |
| rdbsa32_a_cl60 | 32 | 60 mm | 3D-printed polymer case | Omnetics 32-pin |
| rdbsa32_c_cl60 | 32 | 60 mm | 3D-printed titanium housing | Omnetics 32-pin |
| rdbsa32_cc_cl60 | 32 | 60 mm | Epoxy | Omnetics 32-pin |
| rdbsa40_a_cl35 | 40 | 35 mm | 3D-printed polymer case | MRI Omnetics 42-pin |
| rdbsa40_c_cl35 | 40 | 35 mm | 3D-printed titanium housing | MRI Omnetics 42-pin |
| rdbsa40_cc_cl35 | 40 | 35 mm | Epoxy | MRI Omnetics 42-pin |
| rdbsa40_a_cl45 | 40 | 45 mm | 3D-printed polymer case | MRI Omnetics 42-pin |
| rdbsa40_c_cl45 | 40 | 45 mm | 3D-printed titanium housing | MRI Omnetics 42-pin |
| rdbsa40_cc_cl45 | 40 | 45 mm | Epoxy | MRI Omnetics 42-pin |
| rdbsa40_a_cl60 | 40 | 60 mm | 3D-printed polymer case | MRI Omnetics 42-pin |
| rdbsa40_c_cl60 | 40 | 60 mm | 3D-printed titanium housing | MRI Omnetics 42-pin |
| rdbsa40_cc_cl60 | 40 | 60 mm | Epoxy | MRI Omnetics 42-pin |
The 35-60 mm span is what "tens of millimeters" means in practice, and it is set by the target: STN and GP in a large-brain model sit at different depths from the same entry. The lead designs themselves — including the rat-scale variants — are built to order rather than stocked, so the configuration you can have is a conversation with an application scientist, not a catalog lookup. Materials and contact geometry change with the release. Do not rely on a figure quoted in a document.