DBS Series — rDBSA Research Arrays
A DBS Series array (rDBSA, research deep brain stimulation array) reproduces the geometry of a clinical deep-brain lead on a single long shaft with 32 or 40 individually addressable contacts in a line along it, so any contact can stimulate, record, or both.
Because the contacts are addressed individually, the distance between the stimulating contact and the recording contact can be varied within an experiment. The stimulation artifact limits the usable recording window after each pulse, so the recovery interval is generally planned with the stimulation protocol. The stimulator is the Intan RHS2116 in an X-Series stim-record headstage on XDAQ, at 10 nA to 2.5 mA per channel.
Leads are 35, 45, and 60 mm long, for subthalamic nucleus and globus pallidus targets in large-animal models. A rat-sized build keeps the contact arrangement at a smaller scale. The packages are described under DBS package, and the arrays are described in High-Definition Deep Brain Stimulation (HD-DBS) with rDBSA.
The probes can record and stimulate. Research use only.
Typical values
| Catalog designs | 3 |
|---|---|
| Contacts | 32 or 40 along one lead |
| Lead length | 35 to 60 mm |
Related Terms
Cite this term
NeuroNexus. "DBS Series — rDBSA Research Arrays." NeuroNexus Glossary, updated 2026-11-10. https://www.neuronexus.com/resources/glossary/dbs-series