NeuroNexus · neuronexus.com · Glossary
Neural Interface Volume
Neural interface volume is the volume of tissue enclosed by the electrode sites of a design in three dimensions, in cubic millimeters. It is the size of the region with which the design interfaces, and it is compared with the size of the target structure.
It is calculated as the convex hull of the site positions. A single-shank design encloses a thin volume along its span, a multi-shank design encloses a slab the width of its footprint, and a Matrix 3D array encloses a block.
The volume is primarily a tool for comparison. For example, it allows a 64-site single-shank design and a 64-site 4-shank design to be compared by the region that each covers. It is also the denominator of brain coverage density. A volume larger than the target places some sites outside the target. A volume smaller than the target samples part of it, which may be intended (for example, a laminar profile through a cortical column) or may be a limitation. The parameter has no field in a probe name. It is published in the design space and is calculated from the site positions.
Typical values
| Catalog silicon probes | 0.0004 to 4.8 mm³; about 0.05 mm³ is the median |
|---|---|
| Computed as | convex hull of the site positions |
| Design-space parameter | Neural interface volume (interfaceVolMm3) |
Related Terms
Cite this term
NeuroNexus. "Neural Interface Volume." NeuroNexus Glossary, updated 2026-11-10. https://www.neuronexus.com/resources/glossary/neural-interface-volume