NeuroNexus · neuronexus.com · Glossary
Site Span
Electrode site span is the distance along a shank from the first electrode site to the last. It is the length of tissue that one placement samples at once. It is distinct from the shank length, which determines how deep the probe can reach.
Both are stated in millimeters, and the two are often confused. For example, a 10 mm shank with 32 sites at 100 µm has a 3.1 mm span near its tip and about 7 mm of shank without sites behind it.
Site span follows from site count and pitch. For example, 16 sites at 50 µm cover 0.75 mm, 32 sites at 100 µm cover 3.1 mm, and the 3D Matrix arrays reach 10 mm. The span is chosen for the target structure. About 1 mm spans the mouse cortex, about 2 mm spans the rat cortex, and a hippocampal profile that crosses CA1 and the dentate gyrus typically requires 3 mm or more. A span longer than the target places some sites outside it. A span shorter than the target must be positioned on it. The NeuroNexus Vector design uses this approach for deep targets, with a short array on a long support body.
Shank length is determined by the depth of the target plus the tissue above it. It ranges from 1.5 to 60 mm across the catalog, and 5 and 10 mm are the common lengths. A longer shank is less stiff for the same thickness, which is the reason the deepest designs use a 50 µm substrate or a support body. Layouts and their model-number fields are described under Probe Geometry, and targets under Hippocampus.
Typical values
| Span, 16 sites at 50 µm | 0.75 mm |
|---|---|
| Span, 32 sites at 100 µm | 3.1 mm |
| Longest catalog span (Matrix 3D array) | about 10.5 mm |
| Shank length, catalog | 1.5 to 60 mm (5 and 10 mm most common) |
Related Terms
Cite this term
NeuroNexus. "Site Span." NeuroNexus Glossary, updated 2026-11-10. https://www.neuronexus.com/resources/glossary/site-span