NeuroNexus · neuronexus.com · Glossary
Site Count
Site count is the number of electrode sites on one design, which is the number of channels that it can record or stimulate at once.
On a silicon probe it is the product of the number of shanks and the number of sites per shank. On a thin-film grid it is the number of sites on the film. It is a field of the probe name (the 32 in A1x32-5mm-100-177), and the package and headstage are matched to it.
Site count determines coverage, and it does not by itself indicate recording quality. More sites reach more neurons only when the layout places them in additional tissue. More sites in the same tissue record the same neurons from more positions, which is the purpose of a tetrode or high-density layout. Site count is therefore considered together with pitch and span. For example, 32 sites at 100 µm sample about 3 mm of depth, and 32 sites at 25 µm sample 0.8 mm at higher resolution. The site count also determines the compatible hardware. A 16-site or 32-site design mates to a 16-channel or 32-channel connector and headstage, and a 256-site design requires a high-density package. The site count also scales the data volume of each session. Individual sites are described under electrode site.
Typical values
| Catalog silicon probes | 4 to 1,024 sites; 32 is the median |
|---|---|
| Catalog thin-film grids | 3 to 256 sites; 32 is the median |
| Design-space parameter | Site count (numSites) |
Related Terms
Cite this term
NeuroNexus. "Site Count." NeuroNexus Glossary, updated 2026-11-10. https://www.neuronexus.com/resources/glossary/site-count