A SiNAPS probe has an active pixel under each electrode. Amplification, low-pass filtering and 32:1 multiplexing all sit on the shank, so the signal is converted where it is recorded rather than carried out as an analog trace. That is what makes a thousand channels on a 5 mm shank tractable: the cable carries data, not a thousand analog lines. The family is explained once in HD Series ↗.

Research use only. NeuroNexus products and the procedures described here are for non-clinical research use, and must not be used in human or veterinary medical procedures.

The three catalog designs#

Geometry below is derived from the probe descriptors, not quoted from a datasheet.

channelsshanksshank lengthsite pitchactive length
SiNAPS 1S-25625615.5 mm29.6 µm3,754 µm
SiNAPS 4S-10241,0244, at 560 µm5.5 mm29.6 µm3,754 µm per shank
SiNAPS 8S-10241,0248, at 300 µm5.0 mm30.3 µm1,911 µm per shank

Sites are 14 × 14 µm on 50 µm silicon. Every site is read at 20 kHz, and the noise floor is 6.5 µVrms from 300 to 7,500 Hz. Shank separations are center-to-center.

The decision between them#

The two 1,024-channel designs are the same channel budget spent differently. 4S-1024 puts 3,754 µm of active length on each of four shanks 560 µm apart — depth coverage, with lateral sampling wide enough to cross a boundary. 8S-1024 halves the active length to 1,911 µm and puts eight shanks 300 µm apart — a denser lateral map over a shorter depth span.

So the question is not which is higher density; both are 1,024 channels at ~30 µm pitch. It is whether your structure is deep and you need to cross it, or shallow and you need to map across it.

1S-256 is the single-shank case, and it is the one to use when a second shank would be a second penetration you cannot afford.

What active-pixel costs you#

It is record-only. There is no stimulation path through an active-pixel array. A protocol that needs both high site counts and current delivery has to solve that with different hardware — see NeuroNexus Data Acquisition Systems: Platform Overview.

Calibration comes first and is not optional. Unlike a passive array, the per-pixel electronics have to be calibrated before the array means anything. The step is in the SiNAPS Manual.

The data rate is a project constraint. 1,024 channels at 20 kHz is about 41 MB/s, or 1.18 TB in an eight-hour session. Plan storage with the probes — see High-Density Recording: A Planning Guide.

Which experiments SiNAPS is for#

SiNAPS is chosen when the question is a population, not a handful of units: how a thousand neurons across a structure coordinate during a behavior, how activity flows across cortical layers and columns at once, how a circuit's state changes over a session. At about 30 µm pitch every neuron within the detection radius ↗ of the shank appears on several sites, so the spatial oversampling a spike sorter ↗ needs is built into the geometry, and the yield per penetration is set by the tissue rather than by the electrode. The 14 × 14 µm platinum sites are the smallest in the catalog and read as small sites do — local, with a higher impedance and noise floor than a 177 µm² site, which the on-shank amplifier is placed millimeters from the site to manage.

It is not the choice for a study that needs stimulation, a chronic implant where the acquisition path cannot be carried, or a target where a 50 µm substrate and a 5 mm shank do not fit. For those, the passive high-density designs — a 64-site polytrode at 25 µm, an 8-shank Buzsaki at 256 sites — reach a few hundred channels through a conventional package ↗ and headstage, with stimulation and chronic packages available. The comparison across the catalog is in Choosing an Electrode for Your Experiment.

Acquisition#

Smartbox is the acquisition path for SiNAPS, carrying 1,024+ channels over the SiNAPS interface at the probe's fixed 20 kHz. XDAQ with Radiens is the path when SiNAPS runs alongside passive arrays, grids or Neuropixels in the same workspace.

NeuroNexus

Neural probes, data acquisition systems, and analytics software for neuroscience research. Designed and manufactured in Ann Arbor, Michigan.

sales@neuronexus.com
+1 734 913 8858

640 Avis Drive, Suite 200, Ann Arbor, MI 48108

Products

All rights reserved. Copyright © 2026 NeuroNexus.
Policies

Consultation

Cart

History

Resources

Describe your setup and we'll match you to the right probe

Oaks 0.8α

Ctrl+K

Save products while browsing — they'll appear here

Look for the "Add to cart" button on product pages

Log in to see your conversation history and quotes

Sign in

Featured resources

Contextual documents and guides will appear here based on the current page.

Prefer a person? Talk to an Application Scientist