NeuroNexus · neuronexus.com · Glossary
SiNAPS Probe (High-Density Series)
A SiNAPS probe is a CMOS [active-pixel](/resources/glossary/active-pixel-probe) [silicon probe](/resources/glossary/silicon-probe) with an amplifier beneath every electrode site, so it records from all of its sites at once.
SiNAPS stands for Simultaneous Neural Active Pixel Sensor (Angotzi et al. (2019)). Each active pixel amplifies and filters the signal of its site, and the signals are digitized on the probe, so there is no headstage. Every site records the full band, 0.1 Hz to 5 kHz, and no channels are selected before a recording (Corticale). NeuroNexus offers SiNAPS probes as the High-Density Series: 256 sites on one shank, or 1,024 sites on four or eight shanks.
In head-fixed awake mice, the eight-shank design recorded 1,300 single units and mapped current source density across the layers of the hippocampus (Angotzi et al. (2025)). A 1,024-site single-shank design recorded identified neurons in the motor cortex of awake macaques (Angotzi et al. (2026)). A recent review names Neuropixels and SiNAPS probes as the technologies that made recording from thousands of neurons routine (Siegle & Steinmetz (2026)).
SiNAPS probes record only; an active-pixel array has no stimulation path.
NeuroNexus partners with Corticale to commercialize SiNAPS probes. They are packaged as Activus SiNAPS and acquired by the SiNAPS configuration of Smartbox, with Radiens software. Neuropixels, SiNAPS and Silicon Probes: How to Choose compares the options.
Research use only.
Typical values
| Catalog designs | 3 |
|---|---|
| Sites | 256 or 1,024, all recorded simultaneously |
| Shanks | 1, 4, or 8 |
| Shank length | 5 to 5.5 mm |
| Shank width and thickness | 88 µm and 50 µm |
| Site | platinum, 14 × 14 µm (196 µm²) |
| Site pitch | 30 µm, two columns per shank |
| Length of sites per shank | 3,865 µm (1 and 4 shanks), 1,925 µm (8 shanks) |
| Sampling | digitized on the probe, 12-bit, 20 kS/s per site, fixed |
| Bandwidth | full band, 0.1 Hz to 5 kHz, every site |
| Input-referred noise | about 6.5 µV rms, spike band |
| Reference | platinum internal reference on each shank |
| Mode | record only |
References
- Angotzi et al., 2019 — SiNAPS: An implantable active pixel sensor CMOS-probe for simultaneous large-scale neural recordings. Biosens Bioelectron 126, 355–364
- Angotzi et al., 2025 — Multi-shank 1024 channels active SiNAPS probe for large multi-regional topographical electrophysiological mapping of neural dynamics. Adv Sci 12(16), 2416239
- Siegle & Steinmetz, 2026 — Large-scale electrophysiology at single-spike resolution. Nat Rev Neurosci 27, 395–415
- Angotzi et al., 2026 — High-density, identified cell recordings from motor cortex of awake behaving macaques using multiple 1024-channel CMOS probes. IEEE Trans Biomed Eng
- Corticale — SiNAPS technology
Related Terms
Cite this term
NeuroNexus. "SiNAPS Probe (High-Density Series)." NeuroNexus Glossary, updated 2026-11-13. https://www.neuronexus.com/resources/glossary/hd-series