Grid Electrode (Thin-Film Grid)
A grid electrode is a flexible array of electrode sites in a precise grid layout that is placed on a tissue surface, where it records field potentials or delivers stimulation without penetrating the tissue.
It is a surface neural probe, known as an ECoG grid or ECoG array on the cortex and as an EEG grid on the skull, and it is also used on the spinal cord and peripheral nerves. NeuroNexus grid electrodes are thin-film grids, also called polymer electrode grids, made by thin-film microfabrication on a 12 µm polyimide film, in more than 50 catalog designs with 3 to 256 platinum sites, 25 µm to 1 mm in diameter. The same grids are used on the heart.
A thin-film grid is the surface counterpart of the silicon probe. The same thin-film process deposits platinum sites and traces on a polymer film in place of a silicon shank, and the finished array conforms to the surface on which it rests.
A grid on a surface records a different signal from a penetrating probe. The sites are outside the tissue, separated from the neurons by the pia, the dura, or the skull. Individual spikes are therefore generally not recorded, and the array records the field potential, which is the summed activity of the population beneath each site. The same signal is named ECoG on the cortical surface and EEG on the skull (Buzsáki et al. (2012)). Grid sites are correspondingly large, 25 µm to 1 mm across, and the placement determines the signal. An epidural array on intact dura typically records a smaller signal than a subdural array on the pia, which requires opening the dura. A skull-surface array records rhythms and state without a craniotomy. Conformity to a curved or moving surface is an important design requirement, and the NeuroNexus catalog addresses it for each target: accordion and segmented layouts for the beating heart, pre-curved cuffs sized to the diameter of a nerve, and bregma-referenced layouts for the rodent skull.
Four families cover the four surfaces: ECoG grids for the cortex, EEG arrays for the skull, cardiac grids for the epicardium, and nerve cuffs for peripheral nerves. Together they include 53 catalog designs with 3 to 256 sites. Each mounts on a package and mates to the same headstages as a silicon probe. Neural interfaces are not one size fits all; function follows fit. Grid electrodes are used in research and in clinical recording. NeuroNexus grid electrodes are for research use. Research use only.
Typical values
| Catalog designs | 53, in 4 families |
|---|---|
| Substrate | polyimide, 12 µm |
| Sites per array | 3 to 256 |
| Site diameter | 25 µm to 1 mm |
| Site metal | platinum |
| Surfaces | cortex (epidural or subdural), skull, epicardium, peripheral nerve |