Fusion Probes: Double-Sided Microelectrodes
Fusion probes are two catalog probes assembled back to back: twice the active sites, 360° coverage. What it gives and costs, which experiments, how to order.
4 min read
Updated October 12, 2026
NeuroNexus · neuronexus.com · Resources
A Fusion probe has recording sites on both faces of the shank: it is two catalog probes assembled back to back. Two faces double the sites recorded in one penetration, at twice the thickness of a single probe (30 or 100 µm), and the packaging brings both faces out to the headstages. The family is explained once in Fusion Series.
Research use only. NeuroNexus products and the procedures described here are for non-clinical research use, and must not be used in human or veterinary medical procedures.
What the catalog contains#
Eight catalog Fusion packages at 32, 64 and 128 channels, in standard (F) and
headstage-mating (FH) variants, plus wide (W) forms at 32 channels.
Per-design site count, shank length and site geometry are catalog data; read them from the product catalog rather than from this document.
What double-sided buys, and what it costs#
It gives twice the sites per penetration. One insertion records from twice as many sites, which matters most where the structure is small enough that a second penetration is not an option. The assembly is two probes thick, 30 or 100 µm. That added thickness is generally not a significant added tissue displacement with respect to the function of the probe, so the larger, unbiased circumferential recording volume comes at negligible cost.
It costs you a known orientation. With sites on both faces, which face looks at what is set by the rotation of the shank at insertion. That is a protocol step, not a probe property: orientation has to be fixed and recorded at surgery, and it cannot be recovered afterwards from the data alone.
It also costs routing. Both faces reach one connector, which is what the package does, and it is why Fusion designs are specified as a probe-and-package pair rather than a probe alone.
Which experiments a Fusion probe is for#
A Fusion probe is chosen where the target is small enough that a second penetration is not an option and the question needs more units than one face can give: a nucleus a few hundred micrometers across, a thin structure that a multi-shank design would straddle, or a chronic implant where every additional track is a cost the animal pays. Because sites face in two directions, a single shank samples its whole surrounding cylinder rather than a half of it, and a spike sorter sees a neuron on the near face at high amplitude and on the far face at low, which separates units that a one-sided layout would merge. The same doubling applies to stimulation: a site on each face can be driven independently for a 360° field.
It is not the choice where reach matters more than density. The doubled site count arrives as more channels in the same site span, not a longer one; a study that needs a 3 mm depth profile is better served by a linear layout on one face at 100 µm pitch than by two faces at 50 µm, and a study that needs lateral coverage is better served by shanks than by faces.
How a Fusion probe is ordered and read#
Fusion is a build on the catalog families rather than a family of its own: 63 catalog silicon designs are tagged double-sided, across the A-Series, Buzsaki, polytrode and Vector lines, and the Fusion package is what routes both faces to a headstage. The doubled channel count comes out as several 16- or 32-pin Omnetics connectors rather than one larger connector, so a 64-site Fusion probe presents as two 32-channel connectors and needs a headstage or adapter pair to match; the FH builds mate a headstage directly. Confirm the channel map at setup — each face has its own map and the two are folded together through the package — and record the shank's rotation at insertion in the surgical log, because it is the only record of which face looked at what. The package is chosen with Choosing a Probe Package; the design against the target with Choosing an Electrode for Your Experiment.