How an Electrode Records Neural Activity
What an extracellular electrode records, how LFP and spikes differ, why spikes are detected only within about 100 µm, and what impedance and the reference do.
12 min read
Updated November 10, 2026
NeuroNexus · neuronexus.com · Resources
An extracellular electrode records the small voltage that the membrane currents of nearby neurons produce in the tissue around them. A site on a thin-film probe is a metal pad tens of micrometers across. It measures that voltage relative to a reference, and it detects the spikes of neurons within about 150 µm. This document describes what the electrode measures, how spikes and field potentials differ, and how impedance, the reference, and the tissue response affect a recording.
The physiology of extracellular fields is described in detail elsewhere (Buzsáki et al. (2012) ↗). This document covers the parts that affect the choice and use of a probe or a grid, and it links to the glossary for each term.
Research use only. NeuroNexus products and the procedures described here are for non-clinical research use, and must not be used in human or veterinary medical procedures.
What does an extracellular electrode actually measure?#
When a neuron is active, current flows across its membrane and returns through the extracellular space. Because the extracellular space has resistance, these currents produce a voltage in the tissue. An extracellular electrode measures that voltage. All transmembrane currents in the surrounding volume contribute to it, including synaptic currents and action potentials (Buzsáki et al., 2012).
Three points follow for the electrode. First, the signal is small, from tens of microvolts for a spike to about a millivolt for field potentials, and it is recorded against thermal noise, the activity of other neurons, and movement. Second, the measurement is a difference. A site reports the voltage at its position relative to the voltage at a reference, so the reference contributes to every sample. Third, the voltage at a site is the sum of all sources close enough to contribute. Which neurons are recorded depends on where the site is positioned.
On a thin-film probe, the electrode site is a pad of iridium or platinum, with an area of 92 to 1,250 µm² across the NeuroNexus catalog. A thin-film conductor connects it along the shank to the connector at the base. The position of each site relative to every other site is fixed by the mask set, so a laminar profile recorded on one probe has the same geometry on the next probe of that design.
What is the difference between LFP and spikes in an extracellular recording?#
Spikes and the local field potential are parts of the same wide-band voltage on the same site, and they are separated by filtering. A spike is the extracellular transient produced by one neuron's action potential. It lasts under two milliseconds and is typically tens to a few hundred microvolts in amplitude. It is isolated with a band-pass filter, typically about 300 Hz to 6 kHz. The local field potential (LFP) is the part of the signal below about 100 Hz. It is produced by the summed synaptic and membrane currents of many neurons around the site (Buzsáki et al., 2012). The filter settings are choices, and they affect what is counted in each band.
The two signals differ in spatial scale. A spike is recorded only close to the neuron. The LFP spreads through tissue by volume conduction and can include contributions from sources well beyond the range at which spikes are detected, so the term "local" should be interpreted with caution (Buzsáki et al., 2012). A probe may therefore record clear field potentials from a structure in which it does not isolate single units. Site area also affects the two signals differently. On NeuroNexus silicon probes, a relatively small site (177 µm²) is typically used to record the spikes of nearby neurons, and a relatively large site (703 µm²) averages over a larger area and is well suited to field potentials.
The same signal has different names by recording location. It is the LFP in the tissue, the electrocorticogram (ECoG) on the cortical surface, and the EEG at the scalp (Buzsáki et al., 2012). A thin-film grid on the brain surface is separated from the neurons by the pia or the dura, so it generally records field potentials and not the spikes of individual neurons. NeuroNexus grids have site diameters from 25 µm to 1 mm, chosen for the spatial scale of the activity being mapped.
Why is a spike only visible within about a hundred micrometers of a site?#
Spike amplitude decreases steeply with distance, roughly with the square of the distance from the neuron (Buzsáki et al., 2012). Published estimates of the detection radius are similar in scale. Single units can generally be distinguished within about 50 to 100 µm of a site, and spikes can be detected, but not assigned to single units, out to about 150 µm (Quian Quiroga & Panzeri (2009) ↗). The distance is approximate. It depends on the size and morphology of the neuron and on the noise of the site and the recording chain (Siegle & Steinmetz (2026) ↗).
This range informs the choice of site pitch. Sites closer together than the detection radius record the same neuron at different amplitudes, which a spike sorter uses to separate units. In an early study with silicon probes, a spacing under 50 µm was needed to record one neuron on three sites (Drake et al. (1988) ↗). Sites farther apart tend to record different neurons. Tetrode and polytrode layouts place sites within this range for unit isolation, and a linear array with wider spacing provides a continuous profile across depth. The layouts are described under probe geometry.
The same range limits what one shank can record. A shank samples a column of tissue roughly one to three hundred micrometers across, and additional sites within that column record the same neurons again. Recording from more neurons generally requires more shanks or a longer span, which is the purpose of multi-shank and three-dimensional arrays.
What is a good site impedance for a recording electrode?#
Impedance is the opposition that a site presents to current at a given frequency. It is measured at the electrode–tissue interface and is conventionally reported at 1 kHz. There is no single correct value. The appropriate range depends on the site area, the site material, and the amplifier. A 177 µm² iridium site on a NeuroNexus probe typically measures 200 kΩ to 1 MΩ at 1 kHz. Larger sites measure lower, and smaller sites measure higher.
Impedance affects the recorded data in two ways. The interface generates thermal noise in proportion to the square root of its resistance, so a site with higher impedance has a higher noise floor. The site impedance also forms a voltage divider with the input impedance of the amplifier. When the two are comparable, part of the signal is lost, so the amplifier input impedance should be well above the site impedance. For example, the Intan RHD chips in NeuroNexus headstages present 13 MΩ at 1 kHz, which is more than ten times the impedance of a 1 MΩ site.
Impedance can be lowered without enlarging the site. A PEDOT site coating increases the effective surface area within the same footprint, and Z-Coat, the proprietary NeuroNexus form, lowers site impedance by roughly an order of magnitude. SIROF sites measure nominally below 100 kΩ at 177 µm². The benefit in tissue can be smaller than the bench value suggests, because tissue encapsulation adds impedance to every site (Ludwig et al. (2006) ↗). Site material and site area also determine how much charge a site can deliver in stimulation.
Impedance is also a practical check. A site that measures far above its typical value may have a broken trace or a contaminated surface, and a site that measures far below may have a short. Impedance is therefore measured before each session, as described in Silicon Probes: Setup, Handling, Cleaning and Care.
Why is the reference half of every measurement?#
A recording channel reports the difference between two points. One is the site, and the other is the reference electrode. Signals common to both, such as line noise and some movement artifact, are reduced by the subtraction, and signals that differ between them remain. The position of the reference therefore affects every measurement.
A reference near the sites, in quiet tissue or on the probe, cancels the most shared noise and retains the local signal. This placement is typically chosen for spikes. A reference on the skull or in a distant structure cancels less and preserves slow, shared activity such as the field potential across a region. A reference inside the structure being recorded subtracts part of the signal of interest. No physical reference is ideal. A large, low-impedance electrode adds little thermal noise but differs from the sites in impedance and position, so shared noise is not fully cancelled. A single site on the array matches the others but adds its own noise to every channel (Ludwig et al. (2009) ↗). Common average referencing can correct afterwards for part of what the reference did not cancel.
Ground is distinct from the reference. It is the potential that the recording system is built on, tied to the animal so that the amplifier and the tissue share a common potential. On a NeuroNexus probe, the reference and ground are separate wires on the package. Placement is covered in the surgical guides and the EEG referencing guide. Many recordings that appear to have a noisy electrode have a reference problem.
What happens between the site and the file?#
The voltage at the site travels along a thin-film trace on the shank to a bond pad at the base of the probe. The package, which is the interconnect, connector, and covering assembled around the array, routes it from the bond pad to a connector that mates with a headstage. The headstage amplifies the signal and, in current designs, digitizes it at the head, close to the site, which limits the noise that a cable can add. The acquisition system receives the digital stream from each headstage, timestamps it on one clock with stimulation and behavioral events, and writes it to a file. Software reads the file, and it can also read the stream during recording, so that a problem with a site or the reference can be found during the session. Radiens ↗ is the NeuroNexus software for this workflow.
The recording chain is selected starting from the electrode. The probe design determines the package options, the package determines the connector, the connector determines the headstage, and the headstage determines the acquisition system. The selection is described in that order in Choosing an Electrode for Your Experiment.
What does the tissue do to an electrode over days and weeks?#
In an acute recording, the electrode is in tissue for hours and the interface is essentially unchanged. In a chronic recording, the tissue responds to the implant. Over the first weeks, glial cells encapsulate the shank, and the recorded signals change. In one six-week study with silicon probes, site impedance at 1 kHz increased by about 0.5 to 1 MΩ with encapsulation (Ludwig et al., 2006). Some sites lose units and others gain them, so the recorded population changes over time.
Chronic recording therefore depends on materials and packaging as well as geometry. The substrate and its thickness affect how the shank moves against the tissue, and a thinner structure has been associated with less encapsulation (Seymour and Kipke (2007) ↗). The package determines whether the probe is anchored to the skull or floats with the brain. The choice is described in Choosing a Probe Package, and the procedure is described in the chronic surgical guide.
Related documents#
- Neural Probes: What They Are, How They Work, and How to Choose One
- Silicon Probes and Thin-Film Grids: An Introduction
- What Makes a Thin-Film Electrode Good
- Choosing an Electrode for Your Experiment · Choosing a Probe Package
- The Electrode Design Space · Thin-Film Grids for Surface Recording
- Glossary: Extracellular recording · Spike · Local field potential · Detection radius · Impedance · Reference electrode · Signal-to-noise ratio
Citations#
- Drake, K. L., Wise, K. D., Farraye, J., Anderson, D. J., and BeMent, S. L. (1988). Performance of planar multisite microprobes in recording extracellular single-unit intracortical activity. IEEE Transactions on Biomedical Engineering 35(9), 719–732. 10.1109/10.7273 ↗
- Ludwig, K. A., Uram, J. D., Yang, J., Martin, D. C., and Kipke, D. R. (2006). Chronic neural recordings using silicon microelectrode arrays electrochemically deposited with a poly(3,4-ethylenedioxythiophene) (PEDOT) film. Journal of Neural Engineering 3(1), 59–70. 10.1088/1741-2560/3/1/007 ↗
- Seymour, J. P., and Kipke, D. R. (2007). Neural probe design for reduced tissue encapsulation in CNS. Biomaterials 28(25), 3594–3607. 10.1016/j.biomaterials.2007.03.024 ↗
- Quian Quiroga, R., and Panzeri, S. (2009). Extracting information from neuronal populations: information theory and decoding approaches. Nature Reviews Neuroscience 10(3), 173–185. 10.1038/nrn2578 ↗
- Ludwig, K. A., et al. (2009). Using a common average reference to improve cortical neuron recordings from microelectrode arrays. Journal of Neurophysiology 101, 1679–1689. 10.1152/jn.90989.2008 ↗
- Buzsáki, G., Anastassiou, C. A., and Koch, C. (2012). The origin of extracellular fields and currents: EEG, ECoG, LFP and spikes. Nature Reviews Neuroscience 13(6), 407–420. 10.1038/nrn3241 ↗
- Siegle, J. H., and Steinmetz, N. A. (2026). Large-scale electrophysiology at single-spike resolution. Nature Reviews Neuroscience 27, 395–415. 10.1038/s41583-026-01042-4 ↗