NeuroNexus · neuronexus.com · Glossary
A-Series Probes (Buzsaki)
A Buzsáki-layout probe staggers closely spaced sites near the edges of the tip segment of each shank, so neighboring sites record overlapping populations of neurons and a spike sorter separates units by the amplitude pattern across them.
The layout was designed for recording many units from one penetration, in the hippocampal pyramidal layers in particular (Buzsáki (2004)). It is often used in studies of hippocampal replay and spatial navigation. A site spacing under about 50 µm is generally needed to record a neuron with adequate signal-to-noise on several sites at once (Drake et al. (1988)), and spike sorting uses the amplitude differences across those sites to separate units.
Catalog designs have 16 to 256 sites on 1 to 8 shanks that are 1.5 to 15 mm long, with sites of 92 to 177 µm² on 15 or 50 µm silicon. Iridium is the standard site metal. The probes are available on the A-Series, CM, and Hybrid packages.
The probes can record and stimulate. Research use only.
Typical values
| Catalog designs | 16 |
|---|---|
| Sites | 16 to 256 |
| Shanks | 1 to 8 |
| Shank length | 1.5 to 15 mm |
| Site area | 92 to 177 µm² |
| Substrate | 15 or 50 µm silicon |
References
Related Terms
Cite this term
NeuroNexus. "A-Series Probes (Buzsaki)." NeuroNexus Glossary, updated 2026-11-10. https://www.neuronexus.com/resources/glossary/a-series-buzsaki