Skip to content
Inventory sale through Dec 19: B-stock 25–50% off, in-stock A-stock 15% off online.New site. Inventory sale through Dec 19.What's newSee us at SfNSee us at SfN →
Neural Probes · High-Density Series

SiNAPS 4-Shank 1,024-Channel Probe

1,024Sites
4Shanks
5.5mmShank Length
30µmSite Pitch
Part no.SiNAPS-4s-1024-5.5mm
Contents

1,024-site active-pixel silicon probe, 4 shanks

Four mid-length shanks side by side, spread for lateral coverage across columns, with an active pixel beneath every site. Among the 369 probe designs in the design space it sits at the top: 1,024 sites (100th percentile), 5.5 mm shanks (54th percentile of the 366 measured), a 3.77 mm-deep site array (93rd percentile), 158 sites/mm² (43rd percentile) and a 4.75 mm³ neural interface (98th percentile).

Small sites read a local patch at higher impedance, sharpening single-unit isolation, packed densely enough to track a unit across neighbouring sites. The mid-length shanks reach structures a few millimetres down.

Orders as a probe-plus-package assembly, in 2 package configurations. Silicon 50 µm thick. Optical stimulation runs from a separate laser or LED source. Research use only.

Also known as

sinaps_4s_1024

SiNAPS 4-Shank 1,024-Channel ProbeSKU prb_hdd--sinaps_4s_1024_5p5mm

25 specs

Channels1024
Shanks4
Shank length5.5 mm
Shank separation560 µm
Site layoutPixel
Site area196 µm²
Site pitch30 µm
Vertical span3768 µm
SubstrateSilicon
Site materialPlatinum
Thickness50 µm
Neural interface volume4.746 mm³ (higher than 98% of 369 designs)
Site coverage density157.7 sites/mm² (higher than 43% of 369 designs)
Brain coverage density215.8 sites/mm³ (higher than 98% of 369 designs)
ArchitectureActive pixel sensor
In-pixel gain46 dB (DC-4 kHz)
Input noise6.5 µVrms (300-5000 Hz)
Sampling rate20 kSample/s per pixel
Multiplexing32:1
Power<6 µW per electrode-pixel
Electrode pitch30 ± 1 µm
Shank length5656 ± 60 µm
Active length3865 ± 1 µm
Shank width88 ± 2 µm
Shank spacing560 ± 2 µm
SiNAPS 4-Shank 1,024-Channel ProbeSKU prb_hdd--sinaps_4s_1024_5p5mm

Narrow the options with the filters or pick a card, then add the configuration to your cart. When your cart is complete, get pricing, lead time and usage details in one request; we reply within one business day.

2 options

Configure

Compatible Package
All
AVS1024 — 1024-ch Samtec
OAVS1024 — 1024-ch Samtec

Your selection

1,024-site active-pixel silicon probe, 4 shanks, 5.5 mm — 50 μm, AVS1024
AVS1024
Compatible Package: AVS1024 — 1024-ch Samtec
SKUprb_hdd--sinaps_4s_1024_5p5mm.50um.avs1024
An application scientist replies within one business day: fit for your prep, usage details, lead time and pricing.

Options

2 of 2
SiNAPS 4-Shank 1,024-Channel ProbeSKU prb_hdd--sinaps_4s_1024_5p5mm

Active Pixel Sensor

Amplification, filtering and multiplexing sit directly beneath each electrode-pixel — so signal is conditioned and digitized on the shank, not in an external headstage. That is what lets an active-CMOS probe reach thousands of simultaneous channels a passive probe cannot.

46 dB
In-pixel gain
32:1
Multiplexing
<6 µW
Power / pixel
20 kS/s
Per-pixel sampling
6.5 µVrms
Input noise
SiNAPS 4-Shank 1,024-Channel ProbeSKU prb_hdd--sinaps_4s_1024_5p5mm

Loading the compatible products…

SiNAPS 4-Shank 1,024-Channel ProbeSKU prb_hdd--sinaps_4s_1024_5p5mm

14 metrics

Each track spans the smallest to the largest value across catalog designs. The band is the range within its family, and the dot is this design's value. “Top N%” is a rank: how many designs have a larger value.

Sites
1024
41024

Largest among catalog designs · largest in its family

Sites 1024 — higher than 100% of catalog designs and 100% of its family (the largest in its family, 3 designs)
Shanks
4
116

Top 15% of catalog designs

Shanks 4 — higher than 85% of catalog designs and 67% of its family
Shank length
5.5 mm
1 mm60 mm

Top 46% of catalog designs · largest in its family

Shank length 5.5 mm — higher than 54% of catalog designs and 100% of its family (the largest in its family, 3 designs)
Shank separation
0.56 mm
0.05 mm1.8 mm

Top 5% of catalog designs · largest in its family

Shank separation 0.56 mm — higher than 95% of catalog designs and 100% of its family (the largest in its family, 2 designs)
Maximum site depth
5.35 mm
0.972 mm51.29 mm

Top 46% of catalog designs · largest in its family

Maximum site depth 5.35 mm — higher than 54% of catalog designs and 100% of its family (the largest in its family, 3 designs)
Minimum site depth
1.58 mm
0.067 mm44.74 mm

Top 90% of catalog designs

Minimum site depth 1.58 mm — higher than 10% of catalog designs and 67% of its family
Span, x
1.72 mm
0.01 mm6.03 mm

Top 6% of catalog designs

Span, x 1.72 mm — higher than 94% of catalog designs and 67% of its family
Span, y
3.77 mm
0.01 mm10.55 mm

Top 7% of catalog designs · largest in its family

Span, y 3.77 mm — higher than 93% of catalog designs and 100% of its family (the largest in its family, 3 designs)
Site pitch, x
0.03 µm
0.012 µm0.57 µm

Top 29% of catalog designs · largest in its family

Site pitch, x 0.03 µm — higher than 71% of catalog designs and 100% of its family (the largest in its family, 3 designs)
Site pitch, y
0.03 µm
0.01 µm1.5 µm

Top 67% of catalog designs

Site pitch, y 0.03 µm — higher than 33% of catalog designs and 67% of its family
Site outline area
6.5 mm²
0.002 mm²28.74 mm²

Top 2% of catalog designs · largest in its family

Site outline area 6.5 mm² — higher than 98% of catalog designs and 100% of its family (the largest in its family, 3 designs)
Neural interface volume
4.75 mm³
0.149 mm³17.19 mm³

Top 2% of catalog designs · largest in its family

Neural interface volume 4.75 mm³ — higher than 98% of catalog designs and 100% of its family (the largest in its family, 3 designs)
Site coverage density
157.7 sites/mm²
1.66 sites/mm²3047.2 sites/mm²

Top 57% of catalog designs

Site coverage density 157.7 sites/mm² — higher than 43% of catalog designs and 33% of its family
Brain coverage density
215.8 sites/mm³
2.49 sites/mm³393.9 sites/mm³

Top 2% of catalog designs

Brain coverage density 215.8 sites/mm³ — higher than 98% of catalog designs and 33% of its family
SiNAPS 4-Shank 1,024-Channel ProbeSKU prb_hdd--sinaps_4s_1024_5p5mm

6 notes · Choosing between the two 1,024-site designs · What the active-pixel architecture requires · Depth coverage, wide enough to cross a boundary …

SiNAPS 4-Shank 1,024-Channel ProbeSKU prb_hdd--sinaps_4s_1024_5p5mm

6 downloads · 21 links · 3 videos · Choosing an Electrode for Your Experiment · High-Density Recording: A Planning Guide · Neural Recording Chips: A Guide to DAQ ASICs …

SiNAPS 4-Shank 1,024-Channel ProbeSKU prb_hdd--sinaps_4s_1024_5p5mm

Cited in 2,342 publications (product line)

Ask

Cart

History

Resources

Ask us anything about this probe

Products, compatibility, documents and how to choose. Quotes and orders go to our team.

Oaks 0.8α

Ctrl+K

Save products while browsing — they'll appear here

Look for the "Add to cart" button on product pages

Log in to see your conversation history and quotes

Sign in

Resources for this product

Datasheets, application notes, and videos will appear here when connected to the content registry.