SiNAPS 8-Shank 1,024-Channel Probe
1,024-site active-pixel silicon probe, 8 shanks
Eight mid-length shanks side by side, spread for lateral coverage across columns, with an active pixel beneath every site. Among the 369 probe designs in the design space it is in the upper quartile: 1,024 sites (100th percentile), 5 mm shanks (a common choice — 108 of 366), a 1.92 mm-deep site array (80th percentile), 248 sites/mm² (51st percentile) and a 3.2 mm³ neural interface (96th percentile).
Small sites read a local patch at higher impedance, sharpening single-unit isolation, packed densely enough to track a unit across neighbouring sites. The mid-length shanks reach structures a few millimetres down.
Orders as a probe-plus-package assembly, in 2 package configurations. Silicon 50 µm thick. Optical stimulation runs from a separate laser or LED source. Research use only.
sinaps_8s_1024
23 specs
| Channels | 1024 |
|---|---|
| Shanks | 8 |
| Shank length | 5 mm |
| Shank separation | 300 µm |
| Site layout | Pixel |
| Site area | 196 µm² |
| Site pitch | 30 µm |
| Vertical span | 1925 µm |
| Substrate | Silicon |
| Site material | Platinum |
| Thickness | 50 µm |
| Neural interface volume | 3.203 mm³ (higher than 96% of 369 designs) |
| Site coverage density | 248.4 sites/mm² (higher than 51% of 369 designs) |
|---|---|
| Brain coverage density | 319.7 sites/mm³ (higher than 99% of 369 designs) |
| Architecture | Active pixel sensor |
| In-pixel gain | 46 dB (DC-4 kHz) |
| Input noise | 6.5 µVrms (300-5000 Hz) |
| Sampling rate | 20 kSample/s per pixel |
| Multiplexing | 32:1 |
| Power | <6 µW per electrode-pixel |
| Electrode pitch | 30 ± 1 µm |
| Active length | 1925 µm |
| Shank width | 88 ± 2 µm |
Narrow the options with the filters or pick a card, then add the configuration to your cart. When your cart is complete, get pricing, lead time and usage details in one request; we reply within one business day.
2 options
Configure
Your selection
1,024-site active-pixel silicon probe, 8 shanks, 5.0 mm — 50 μm, AVS1024
AVS1024Options
2 of 2AVS1024 — 1024-ch Samtec
OAVS1024 — 1024-ch Samtec
Amplification, filtering and multiplexing sit directly beneath each electrode-pixel — so signal is conditioned and digitized on the shank, not in an external headstage. That is what lets an active-CMOS probe reach thousands of simultaneous channels a passive probe cannot.
46 dB
In-pixel gain32:1
Multiplexing<6 µW
Power / pixel20 kS/s
Per-pixel sampling6.5 µVrms
Input noiseLoading the compatible products…
14 metrics
Each track spans the smallest to the largest value across catalog designs. The band is the range within its family, and the dot is this design's value. “Top N%” is a rank: how many designs have a larger value.
Largest among catalog designs · largest in its family
Sites 1024 — higher than 100% of catalog designs and 100% of its family (the largest in its family, 3 designs)Top 3% of catalog designs · largest in its family
Shanks 8 — higher than 97% of catalog designs and 100% of its family (the largest in its family, 3 designs)Top 48% of catalog designs
Shank length 5 mm — higher than 52% of catalog designs and 33% of its familyTop 24% of catalog designs
Shank separation 0.3 mm — higher than 76% of catalog designs and 50% of its familyTop 77% of catalog designs
Maximum site depth 4.81 mm — higher than 23% of catalog designs and 33% of its familyTop 73% of catalog designs · largest in its family
Minimum site depth 2.89 mm — higher than 27% of catalog designs and 100% of its family (the largest in its family, 3 designs)Top 3% of catalog designs · largest in its family
Span, x 2.14 mm — higher than 97% of catalog designs and 100% of its family (the largest in its family, 3 designs)Top 20% of catalog designs
Span, y 1.92 mm — higher than 80% of catalog designs and 33% of its familyTop 30% of catalog designs
Site pitch, x 0.028 µm — higher than 70% of catalog designs and 33% of its familyTop 65% of catalog designs · largest in its family
Site pitch, y 0.03 µm — higher than 35% of catalog designs and 100% of its family (the largest in its family, 3 designs)Top 4% of catalog designs
Site outline area 4.12 mm² — higher than 96% of catalog designs and 67% of its familyTop 4% of catalog designs
Neural interface volume 3.2 mm³ — higher than 96% of catalog designs and 67% of its familyTop 49% of catalog designs
Site coverage density 248.4 sites/mm² — higher than 51% of catalog designs and 67% of its familyLargest among catalog designs · largest in its family
Brain coverage density 319.7 sites/mm³ — higher than 100% of catalog designs and 100% of its family (the largest in its family, 3 designs)6 notes · Choosing between the two 1,024-site designs · What the active-pixel architecture requires · A denser lateral map over a shorter span …
6 downloads · 21 links · 3 videos · High-Density Recording: A Planning Guide · SiNAPS: Active High-Density Microelectrodes · SiNAPS Probe Maps …
Cited in 2,342 publications (product line)